Transactions of the Society of Instrument and Control Engineers
Online ISSN : 1883-8189
Print ISSN : 0453-4654
ISSN-L : 0453-4654
Paper
Comparison of Installability and Robustness in Wearable Tele-echography Device with Body Attachment Mechanisms
Tomofumi ASAYAMAKeiichiro ITOKoichi TSURUTAShigeki SUGANOKyota NAKAMURAHiroyasu IWATA
Author information
JOURNAL FREE ACCESS

2013 Volume 49 Issue 1 Pages 86-92

Details
Abstract
It is necessary for the patient who met with a traffic accident to perform echography immediately. We have developed the tele-echography device which the doctor who is in the hospital during transportation can inspect by remote control. In this paper, we report new attachment mechanisms of the device. Conventional system has an attachment mechanism with a corset. This mechanism has to pass a corset through between patient's body and ground. It is hard to attach this device to patient used by this method. Furthermore patient suffers damage by moving their body. We developed, therefore, new mechanisms which we can put it on the front of the body and evaluated their installability and robustness in a vehicle.
Content from these authors
© 2013 The Society of Instrument and Control Engineers
Previous article Next article
feedback
Top