計測自動制御学会論文集
Online ISSN : 1883-8189
Print ISSN : 0453-4654
ISSN-L : 0453-4654
相関の微分を用いる動特性測定法
柏木 濶灘吉 一隆
著者情報
ジャーナル フリー

1971 年 7 巻 1 号 p. 39-46

詳細
抄録
When the autocorrelation function of an input signal to a system is Dirac's delta (δ-) function, the impulse response of the system is obtained as the crosscorrelation function between the input and the output of the system. In practice, however, there exist some cases where such an input signal cannot be used because of its lack of power, or because of poor frequency characteristics of the devices actually applying the input. In these cases the impulse response is not obtained simply from the crosscorrelation, but indirectly by a troublesome deconvolution. In the present paper a simple deconvolution method is proposed in which the impulse response is determined from a linear combination of the derivatives of the crosscorrelation function. The method is particularly simple when the autocorrelation function φxx(τ) is given as φxx(τ)=e-λ|τ| or φxx(τ)={1-|τ|/T(|τ|≤T) 0(|τ|>T)
This method does not require the function φxx(τ) to be the δ-function, but to be not differentiable at some τ's. Essentially, the method utilizes the δ-functions appearing when φxx(τ) is differentiated in the sense of the distribution proposed by L. Schwartz. The experimental values obtained by the computer simulation agree well with the theoretical ones, and the errors estimated are negligibly small.
著者関連情報
© 社団法人 計測自動制御学会
前の記事 次の記事
feedback
Top