A Determination Method of the Femoral Bone Coordinate System by Integrating an Ultrasound Device with a 3D Electromagnetic Device
*Kouki NagamuneHirotsugu MuratsuKoji NishimotoYuichi HoshinoKoichi TanakaKiyonori MizunoMasayoshi YagiRyosuke KurodaShinichi YoshiyaMasahiro Kurosaka
Author information
CONFERENCE PROCEEDINGS FREE ACCESS

Details
Abstract
This paper proposes new determination method of the bone axis of the femur by integrating an ultrasound device with a 3D electromagnetic device. Recently, quantitative measurement method intra-surgery has been received grate attention. Problems how to determine the bone axes are always raised in the clinical situation. Especially, the conventional method usually uses the major trochanter as a reference point, however, it cannot be access in anterior cruciate ligament reconstruction. This study provides one of the solutions about the problem by integrating an ultrasound device with a 3D electromagnetic device. In the experiment, our method was compared with the conventional method where both of methods were applied for one cadaveric body. As a result, the 6DOF calculated by our method without using the major trochanter, was similar to the conventional method.