Hideki ShigematsuGo YoshidaHiroki UshirozakoKenta KurosuNobuaki TadokoroMasahiro FunabaShigenori KawabataJun HashimotoMuneharu AndoShinichirou TaniguchiMasahito TakahashiNaoki SegiHiroaki NakashimaShiro ImagamaShinji MoritoKei YamadaTsunenori TakataniTsukasa KanchikuYasushi FujiwaraHiroshi IwasakiKanichiro WadaNaoya YamamotoKazuyoshi KobayashiAkimasa YasudaKazuyoshi NakanishiYasuhito TanakaYukihiro MatsuyamaKatsushi Takeshita
Author information
JOURNAL OPEN ACCESS

2025 Volume 9 Issue 2 Pages 173-178

Details
Abstract

Introduction: Multimodal intraoperative neurophysiological monitoring (IONM) -such as monitoring muscle-evoked potentials after transcranial electrical stimulation (Tc-MEP) with somatosensory-evoked potential (SEP) after electrical stimulation of the peripheral nerve-is recommended in spine surgeries to prevent iatrogenic neurological complications. However, the effect of using Tc-MEP with SEP to protect against neurological complications, particularly motor function, remains unknown. In clinical settings, changes due to Tc-MEP meeting the alarm points must be a potential neurological injury. This retrospective study, focusing on true-positive (TP) cases, aimed to clarify the change in the SEP waveform simultaneously with the Tc-MEP alarm.

Methods: We included 68 patients with TP who had Tc-MEP changes and new postoperative motor weakness at more than one level of the manual muscle test after surgery. We compared the cases based on the category of spine surgery and paralysis type. We evaluated sex, age at spine surgery (high- or non high-risk), and paralysis type (segmental, long tract, or both). We defined the alarm points as follows: >70% decrease in Tc-MEP wave amplitudes, >50% decrease in wave amplitudes, or 10% extension of SEP latency. Next, we evaluated the SEP wave changes with a Tc-MEP alarm.

Results: All patients showed progressive motor weakness after surgery, and 21 patients (31%) showed SEP changes at the same time as the Tc-MEP alarm. There were no statistically significant differences in the ratio of SEP change between the two groups according to the spine surgery category or among the three groups according to the paralysis type.

Conclusions: Multimodal IONM is an important tool. However, the SEP changes do not necessarily appear immediately after the Tc-MEP alarm. Spine surgeons should appropriately treat Tc-MEP alarms to preserve motor function, regardless of SEP changes.

Fullsize Image
Content from these authors
© 2025 The Japanese Society for Spine Surgery and Related Research.

SSRR is an Open Access journal distributed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License. Anyone may download, reuse, copy, reprint, or distribute articles published in the journal for not-for-profit purposes if they cite the original authors and source properly. If you remix, transform, or build upon the material, you may not distribute the modified material.