鉄と鋼
Online ISSN : 1883-2954
Print ISSN : 0021-1575
ISSN-L : 0021-1575
X線マイクロアナライザーによるCr-Ni-Moオーステナイト鋼における固溶窒素の定量分析
小野 長門田島 至澤田 滋梶原 正憲菊池 實
著者情報
ジャーナル フリー

1992 年 78 巻 1 号 p. 178-185

詳細
抄録

Five 2025Cr-Ni-Mo austenitic steels were prepared with a nickel concentration range between 5 and 28 mass% and a molybdenum concentration range between zero and 4 mass%. Five series of Cr-Ni-Mo austenitic steels with different nitrogen concentrations ranging from 0.01 to 0.76 mass% were prepared by nitrogenizing under different nitrogen gas pressures at 1 523 K without changing the matrix composition. They were used to make calibration curves between the nitrogen concentrations and NKα intensities. A coated multi-layer X-ray dispersion element, specifically designed for measuring NKα intensities, was used as an analyzing crystal. Correlation between the nitrogen concentrations and the NKα intensities was represented by a straight line. Gradients and background intensities for five calibration curves are all the same within the present experimental accuracy. Nitrogen partitioning between ferrite and austenite in duplex stainless steels was determined by utilizing the calibration curves for five Cr-Ni-Mo austenitic steels. Nitrogen in solution both in ferrite and austenite was analyzed with an accuracy of ±0.02 mass%.

著者関連情報
© 一般社団法人 日本鉄鋼協会
前の記事 次の記事
feedback
Top