2025 年 E108.B 巻 9 号 p. 974-982
This paper is intended to help provide an introduction for EMC analysis, design and signal integrity assurance for wearable wireless biosignal sensors. First, a mechanism by which external electromagnetic field couple to the biosignal sensor circuit is explained in terms of common-mode to differential-mode conversion. Based on this, some noise quantification and suppression techniques for biosignal sensors are introduced. Next, focusing on reducing inter-symbol interference in wideband communications from the surface to the inside of the human body, a pre-emphasis technique which guarantee a balance between signal integrity improvement and signal-to-noise power degradation is proposed and proved useful.