ACTA HISTOCHEMICA ET CYTOCHEMICA
Online ISSN : 1347-5800
Print ISSN : 0044-5991
ISSN-L : 0044-5991

This article has now been updated. Please use the final version.

A Device for Ribbon Collection for Array Tomography with Scanning Electron Microscopy
Taro KoikeYosky KataokaMitsuyo MaedaYuji HasebeYuuki YamaguchiMitsuo SugaAkira SaitoHisao Yamada
Author information
JOURNAL FREE ACCESS Advance online publication

Article ID: 17013

Details
Abstract

“Array tomography” is a method used to observe the fine structure of cells and tissues in a three-dimensional view. In this method, serial ultrathin sections in the ribbon state (ribbons) are mounted on a solid substrate and observed by scanning electron microscopy (SEM). The method may also be used in conjunction with post-embedding immunocytochemistry. However, it is difficult to mount many serial ribbons on a substrate manually. We developed an inexpensive laboratory-made device that mounts ribbons by pulling a nylon fishing line and lifting the substrate up from the water in a knife boat. Using this device, we succeeded in mounting several ribbons consisting a mean of 205.6 (SD: 37.7) serial ultrathin sections on 1.25 (SD: 0.06) × 1.25 (SD: 0.06)-cm silicon substrates. Furthermore, it was confirmed that our method is suitable for ribbons derived from water-soluble resin blocks. We were also able to stain the specimens by post-embedding immunocytochemistry. Thus, our method is useful in mounting numerus sections on a substrate for array tomography with SEM.

Content from these authors
© 2017 The Japan Society of Histochemistry and Cytochemistry
feedback
Top