Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Cryogen-Free Automated Gas Chromatograph System for Monitoring of Halocarbons in the Atmosphere at Background Concentration Levels
Sakae TOYODATakeshi TOMINAGAYoshihiro MAKIDE
著者情報
ジャーナル フリー

1998 年 14 巻 5 号 p. 917-923

詳細
抄録
An automated gas chromatograph/electron capture detector (GC/ECD) system is developed for the monitoring of major seven halocarbons in the atmosphere (CCl2F2: CFC-12, CCl3F: CFC-11, CCl2FCClF2: CFC-113, CH3CCl3, CCl4, CHCl=CCl2, and CCl2=CCl2) at the tropospheric background concentration levels. By using two columns in series for each of two groups divided by the volatility of the compounds, the system requires neither cryogen for preconcentration nor any physical/chemical dryer for water removing. A sample amount of ca.18 ml STP is sufficient to detect these species, and analysis of sub-ambient atmospheric-pressure sample in a canister is also possible. Monitoring of these halo-carbon concentrations was carried out at 90-min intervals in the urban area.
著者関連情報
© 1998 by The Japan Society for Analytical Chemistry
前の記事 次の記事