Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Sensitive and Simple Determination of Diclofenac Sodium by Use of a Continuous Flow-Injection Procedure with Solid-Phase Spectroscopic Detection
P. ORTEGA-BARRALESA. RUIZ-MEDINAM. L. FERNÁNDEZ-DE CÓRDOVAA. MOLINA-DIAZ
Author information
JOURNAL FREE ACCESS

1999 Volume 15 Issue 10 Pages 985-989

Details
Abstract
A flow-through sensor for the determination of diclofenac sodium was developed, based on retention of the analyte on a Sephadex QAE A-25 anion-exchange resin packed in a flow-cell of 1 mm of optical path length, and monitoring of its intrinsic absorbance by UV-spectrophotometry at 281 nm. Diclofenac could be determined in the concentration ranges 2.0 - 40.0, 1.0 - 22.0 and 0.5 - 14.0 µg ml-1 with RSD (%) ranging from 1.05 to 1.53 for sample volumes of 300, 600 and 1200 µl, respectively. The proposed sensor was satisfactorily applied to the rapid determination of diclofenac in commercial pharmaceutical preparations and in semi-synthetic pharmaceuticals containing diclofenac and paracetamol.
Content from these authors
© 1999 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top