Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Application of Sequential Injection Analysis to the Assay of Lead Retention Characteristics by Poly(vinylpyrrolidone): Trace Analysis of Lead in Waters
Alberto N. ARAÚJORui Cerdeira C. COSTAJosé L. F. C. LIMA
Author information
JOURNAL FREE ACCESS

1999 Volume 15 Issue 10 Pages 991-994

Details
Abstract
This paper presents the use of poly(vinylpyrrolidone) (PVP), which is a low-cost polymer for lead preconcentration in natural-water samples. The weak granular structure of this polymer imposed the use of a stainless-steel filter holder for its conditioning in a thin filtering layer. This unit was inserted into a sequential injection analysis system coupled to a flame atomic absorption spectrophotometer (SIA/FAAS). The developed system allows sampling rates of 16 samples/h with a 60-fold enrichment factor and a detection limit of less than 5 µg/l for a 13 ml sample volume. The performance of the present system was also assessed by carrying out recovery trials, which provided results of between 96% and 101%. The precision of the methodology developed has never been over 3% regarding RSD and 5% for the deviations to the conventional procedure.
Content from these authors
© 1999 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top