Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Implementation of a Generalized Standard Addition Method in a Flow Injection System Using Merging-Zones and Gradient Exploitation
Edvan C. SILVAValdomiro L. MARTINS Valdomiro L. MARTINSAlessandro F. ARAÚJOMário César U. ARAÚJO
Author information
JOURNAL FREE ACCESS

1999 Volume 15 Issue 12 Pages 1235-1240

Details
Abstract
A novel strategy for implementing a generalized standard addition method (GSAM) in a simple two-channel flow injection (FI) system and for correcting matrix effects and spectral interferences in multicomponent analysis is described. The FI-GSAM system exploits the concentration gradients of the sample and the standard solution, requires only one standard solution per analyte and provides several addition levels using a single simultaneous injection of the sample and of each standard solution. Because the matrix effects change at every addition level, a novel mathematical model for this FIGSAM has been derived. The method was critically evaluated for simultaneous K, Ca and Na determinations by flame photometry in synthetic samples with different ethanol or glycerol contents, added to the samples to promote matrix effects. The results show good agreement with the expected values, with the relative error and the overall relative standard deviation usually being <5% (n=5) for 2.5 - 5.0 Na, 20.0 K and 1000 mg l-1 Ca. The elapsed time for the simultaneous determination of three analytes was about 2 min, consuming 400 μl of the sample and 100 μl of each standard solution.
Content from these authors
© 1999 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top