Experimental Optimization of Selective Hydrazine Detection in Flow Injection Analysis Using a Carbon Paste Electrode Modified with Copper Porphyrin Occluded into Zeolite Cavity
Silvia V. GUERRALauro T. KUBOTAClaudia R. XAVIERShirley NAKAGAKI
Author information
JOURNAL FREE ACCESS

1999 Volume 15 Issue 12 Pages 1231-1234

Details
Abstract
A carbon paste electrode modified with copper porphyrin occluded into zeolite cavity was developed for sensing hydrazine in flow system. The sensor presented a high sensitivity and selectivity for hydrazine concentration in the flow system. The flow system was optimized by factorial design considering a two level, three factor factorial design (the factors were applied potential, sampling volume and flow rate). The sensitivity and analytical frequency were considered as the most important parameters for the system. The best condition for the system was sampling volume of 50 μl, applied potential of 270 mV vs. SCE and a flow rate of 1 ml min-1. In this optimized condition, the electrode can improve its stability, sensitivity and selectivity. The operational range was between 5 up to 60 μmol l-1, adjusted by the equation Y=1.9(±0.1)+0.20(±0.01)[hydrazine]-0.0014(±0.0001)[hydrazine]2, with a correlation coefficient of 0.9995 for n=8. The analytical frequency was about 70 determinations per hour and the system was stable for at least two months in continuous use.