Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Powder Microelectrodes for Capillary Electrophoresis with Amperometric Detection
Shen HUMin ZHANGDai-Wen PANGJie-Ke CHENG
著者情報
ジャーナル フリー

2000 年 16 巻 8 号 p. 807-810

詳細
抄録
In this paper, we first report the use of powder microelectrodes (PMEs) as the amperometric detector for thiols, hydrazine and hydroxylamine in capillary electrophoresis (CE). This PME, with graphite-supported metalloporphyrin powder as catalyst, offers a few advantages as an amperometric detector in CE. Concentration limit of detection and mass limit of detection obtained for L-cysteine at the PME could reach 88 nM and 230 amol, respectively. The response of thiols at the PME was very reproducible since the electrode fouling was reduced. A relative standard deviation of peak heights (n = 8) of 1.8% and a relative standard deviation of migration time (n = 8) of 1.0% can be obtained for L-cysteine. The material packed in the PME can easily be changed. Hence, it is very convenient to choose suitable catalysts to detect special analytes, and this specially designed microelectrode should be applicable to many other compounds of biological interest.
著者関連情報
© 2000 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top