Application of an HPLC-FTIR Modified Thermospray Interface for Analysis of Dye Samples
Mohammad A MOTTALEBDavid LITTLEJOHN
著者情報
ジャーナル フリー

2001 年 17 巻 3 号 p. 429-434

詳細
抄録
Previously, we developed a reversed-phase HPLC method compatible to high performance liquid chromatography diffuse reflectance Fourier-transform infrared (HPLC-FTIR) thermospray interface for the analysis of dyes. Dye separation achieved with a mixed-mode (SCX-ODS) column using a small gradient (90 to 80% water with acetic acid) and pH 3.25; 10 to 20% acetonitrile was considered to be suitable for HPLC-FTIR. A constant-voltage setting for the thermospray temperature (227°C) was successfully used for this gradient condition. The HPLC-separated components deposited as a series of concentrated spots on a moving tape were scanned by specially developed HPLC-FTIR software. Excellent repeatability of the thermospray deposition FTIR chromatograms and IR spectra was obtained. The interface-derived spectra of the separated components of formulated and purified reactive dyes were compared and differences in spectral features were observed.
著者関連情報
© 2001 by The Japan Society for Analytical Chemistry