Construction and Performance Characteristics of New Fentanyl-Selective Plastic Membrane Electrode
Li Juan PENGMeng Liang WENYun YAO
著者情報
ジャーナル フリー

2001 年 17 巻 7 号 p. 815-818

詳細
抄録

The construction and performance characteristics of an ion-selective electrode for fentanyl-drug cation, based on an ion-pair complex with tetrakis[3,5-bis-(trifluoromethyl)phenyl]borate anion in a PVC matrix were studied. A linear response for 1 × 10-5 mol dm-3 to 1 × 10-2 mol dm-3 drug with a slope of 57.9 ± 0.5 mV/decade was established. The optimum pH range was 2 to 6. The lower detection limit was 6.29 × 10-6 mol dm-3 fentanyl citrate (2.1165 µg cm-3 fentanyl). There were negligible interferences from a number of inorganic cations, structural analogues, and some common drug additives in injections. The electrode proposed has been successfully applied to determine fentanyl citrate in injections. The results correlated well with those obtained by the United States Pharmacopoeia standard procedure.

著者関連情報
© 2001 by The Japan Society for Analytical Chemistry