Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Glucose Sensing Based on Interdigitated Array Microelectrode
Peng JINAkira YAMAGUCHIFumika ASARI OIShigeki MATSUOJiubin TANHiroaki MISAWA
著者情報
ジャーナル フリー

2001 年 17 巻 7 号 p. 841-846

詳細
抄録

A micro glucose sensor consisting of an interdigitated array gold microelectrode was developed. The interdigitated array structure, which has 10 µm band width and 10 µm band gap, was fabricated in a small region (2.5 × 5 mm2) on a quartz substrate. Glucose oxidase was chemically fixed onto the electrode surface through self-assembled monolayer of 11-mercaptoundecanoic acid; ferroceneacetic acid was used as electron mediator. Electrochemical properties of the glucose oxidase-immobilized microelectrode were investigated by cyclic voltammogram measurements. Results confirmed that the reductive ferroceneacetic acid generated at counter electrode diffuses through a narrow band gap (10 µm) and can reach the working electrode surface.

著者関連情報
© 2001 by The Japan Society for Analytical Chemistry