Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Capillary Electrochromatography on Monolithic Silica Columns
Hiroshi KOBAYASHIChris SMITHKen HOSOYATohru IKEGAMINobuo TANAKA
著者情報
ジャーナル フリー

2002 年 18 巻 1 号 p. 89-92

詳細
抄録
Monolithic silica columns prepared from tetramethoxysilane in a fused-silica capillary was octadecylsilylated, and evaluated by capillary electrochromatography. A plate height of 3 - 4 µm was obtained in an 80% acetonitrile-20% aqueous buffer (pH 8) under optimized conditions. An ODS phase having a high surface coverage, prepared by using octadecyldimethylchlorosilane, resulted in slow separation and a low column efficiency due to a slow electroosmotic flow. The results show that faster separation is achieved (i) with an ODS phase with a low surface coverage, (ii) with an ODS phase prepared by using octadecyltrichlorosilane, or (iii) by pressure-assisted CEC utilizing the high permeability of monolithic silica columns.
著者関連情報
© 2002 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top