Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Flow Injection/Atomic Absorption Spectrometric Determination of Zineb in Commercial Formulations of Pesticide Based on Slurry Sampling
Ricardo J. CASSELLAVerônica A. SALIMSalvador GARRIGUESRicardo E. SANTELLIMiguel de la GUARDIA
著者情報
ジャーナル フリー

2002 年 18 巻 11 号 p. 1253-1256

詳細
抄録
This paper reports on a new strategy for the slurry sampling determination of dithiocarbamate pesticide zineb {[ethylenebis(dithiocarbamato)]zinc} employing a FIA system with a flame atomic absorption spectrometry detector. In the flow system, an on-line alkaline hydrolysis of the pesticide is performed, allowing the release of Zn(II) ions to the solution, which are easily detected by a flame AAS technique. Several parameters that could affect the performance of the analytical methodology were studied, such as the concentration of NH3(aq) used in the hydrolysis step, the effect of the presence of Triton X-100 on the sensitivity and precision, and the FIA parameters (carrier flow rate and mixing coil volume). Under optimized conditions, aqueous slurries containing 2.5 to 25 μg ml-1 zineb provided good linear calibration fits. From the obtained data, a detection limit (3σ) of 1.0 μg ml-1 zineb was found and a repeatability of 2.7% was obtained from 12 measurements of a slurry containing 2.5 μg ml-1 zineb. On the other hand, a precision (reproducibility) of 7.8% was achieved from three determinations of a sample containing 128 mg g-1 of the pesticide. Also, the developed system provides a sampling frequency of 72 h-1.
著者関連情報
© 2002 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top