Abstract
A rapid and reproducible method is described that employs solid-phase extraction (SPE) using dichloromethane, followed by gas chromatography (GC) with flame ionization detection for the determination of benzene, toluene, ethylbenzene, xylene and cumene (BTEXC) from Buriganga River water of Bangladesh. The method was applied to detect BTEXC in a sample collected from the surface, or 5 cm depth of water. Two-hundred milliliters of n-hexane-pretreated and filtered water samples were applied directly to a C18 SPE column. BTEXC were extracted with dichloromethane and the BTEX concentrations were obtained to be 0.1 to 0.37 μg ml-1. The highest concentration of benzene was found as 0.37 μg ml-1 with a relative standard deviation (RSD) of 6.2%; cumene was not detected. The factors influencing SPE e.g., adsorbent types, sample load volume, eluting solvent, headspace and temperatures, were investigated. A cartridge containing a C18 adsorbent and using dichloromethane gave a better performance for the extraction of BTEXC from water. Average recoveries exceeding 90% could be achieved for cumene at 4°C with a 2.7% RSD.