Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Development of a Compact Supersonic Jet/Multiphoton Ionization/Time-of-Flight Mass Spectrometer for the On-site Analysis of Dioxin, Part I: Evaluation of Basic Performance
Junichi MATSUMOTOBunji NAKANOTotaro IMASAKA
Author information
JOURNAL FREE ACCESS

2003 Volume 19 Issue 3 Pages 379-382

Details
Abstract
A new type of compact supersonic jet/resonance-enhanced multiphoton ionization/time-of-flight mass spectrometer is described. The analytical instrucment, consisting of a single turbo molecular pump equipped with a rotary pump, was maintained at <2×10-3 Pa when a 0.3-atm sample was injected into a vacuum at 10-Hz using a 200-μs pulse valve. The diameters of the extraction and ground skimmer electrodes were expanded to 30 mm in order to avoid strong focusing and defocusing of the ion, and the optimum conditions for the system were investigated. The mass spectrometer functioned as expected: (1) no defocusing of the ion beam was observed even when the potential of the einzel lens was adjusted to zero; (2) the direction of the ion beam to an assembly of microchannel plates deviated in the expected manner when the potential of the defection electrode was changed from 0 to 30 V.
Content from these authors
© 2003 by The Japan Society for Analytical Chemistry
Previous article Next article