Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
A Miniaturized FIA System for the Determination of Residual Chlorine in Environmental Water Samples
Jiye JINYasutada SUZUKINaoyoshi ISHIKAWAToyohide TAKEUCHI
著者情報
ジャーナル フリー

2004 年 20 巻 1 号 p. 205-207

詳細
抄録
The present study was aimed at developing a portable miniaturized FIA system with a two-electrode based amperometric detector for on-site monitoring of residual chlorine in environmental samples. The electrode potential control and current data collection were conducted with a home-designed potentiostat with a size of 150 × 100 × 40 mm and the weight of 350 g. It was confirmed that the amperometric response increased in proportion to the chlorine concentration up to 5 μg/ml, and the detection limit for chlorine was 0.05 μg/ml. Besides a drastic reduction in the size of the analytical system, such miniaturization achieved the advantages of increased speed, minimal mobile-phase consumption, and high sensitivity. The proposed method was successfully applied to the determination of residual chlorine in samples of pool water and tap water. The results were well agreed with those obtained by the N,N-diethyl-p-phenylenediamine (DPD) photometric method.
著者関連情報
© 2004 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top