Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Large-Volume Injection Electrothermal AAS Combined with Tungsten-treated Pyrolytic Graphite Furnace: Determination of Cadmium in Tap, Snow and River Water Samples with Phosphate Modifier
Shoji IMAISatoshi ISHIKAWAYoichi KIKUCHIAkira YONETANI
著者情報
ジャーナル フリー

2004 年 20 巻 3 号 p. 575-577

詳細
抄録

A large-volume (100 µl) injection-ETAAS with W-treated PG furnace combined with a phosphate modifier was applied to the determination of unpolluted levels of Cd in tap, snow and river-water samples. The limit of detection of 1.1 ng l-1 was observed for a 4 w/v% NH4H2PO4 modifer. Matrix interference studies were tested for major ion species well found in fresh water. The direct determination of Cd in certified river water (12 ± 2 ng l-1) was carried out and the observed value was in agreement with the certified one. The good recoveries of Cd added to real environmental water samples were also observed. This method was applied to the determination of Cd in unpolluted environmental water samples.

著者関連情報
© 2004 by The Japan Society for Analytical Chemistry