Determination of Ambroxol in an Automated Multi-Pumping Pulsed Flow System
João L. M. SANTOSArnaud CLAUSSEJosé L. F. C. LIMAM. L. M. F. S. SARAIVAAntónio O. S. RANGEL
Author information
JOURNAL FREE ACCESS

2005 Volume 21 Issue 4 Pages 461-464

Details
Abstract
A new flow methodology exploiting the multi-pumping approach was developed for the spectrophotometric determination of ambroxol hydrochloride in pharmaceutical preparations. The flow manifold was implemented by using, exclusively, multiple solenoid-actuated micro-pumps, which acted simultaneously as sample insertion, solutions propelling and reagents commutation units. Linear calibration plots were obtained over an ambroxol concentration ranging from 10 to 200 mg l-1 (r.s.d. < 0.5%, n = 15) and a sampling rate of about 60 samples per hour (flow rate = 1.92 ml min-1, sample volume = 80 µl).