Electroanalysis of Tetracycline Using Nickel-implanted Boron-doped Diamond Thin Film Electrode Applied to Flow Injection System
Surudee TREETEPVIJITSuchada CHUANUWATANAKULYasuaki EINAGARika SATOOrawon CHAILAPAKUL
著者情報
ジャーナル フリー

2005 年 21 巻 5 号 p. 531-535

詳細
抄録
The electrochemical analysis of tetracycline was investigated using nickel-implanted boron-doped diamond thin film electrode by cyclic voltammetry and amperometry with a flow injection system. Cyclic voltammetry was used to study the electrochemical oxidation of tetracycline. Comparison experiments were carried out using as-deposited boron-doped diamond thin film electrode (BDD). Nickel-implanted boron-doped diamond thin film electrode (Ni-DIA) provided well-resolved oxidation irreversible cyclic voltammograms. The current signals were higher than those obtained using the as-deposited BDD electrode. Results using nickel-implanted boron-doped diamond thin film electrode in flow injection system coupled with amperometric detection are presented. The optimum potential for tetracycline was 1.55 V versus Ag/AgCl. The linear range of 1.0 to 100 µM and the detection limit of 10 nM were obtained. In addition, the application for drug formulation was also investigated.