Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Simultaneous Determination of Cobalt, Copper and Zinc by Energy Dispersive X-ray Fluorescence Spectrometry after Preconcentration on PAR-loaded Ion-Exchange Resin
Zi-Tao JIANGJimmy C. YUHo-Yan LIU
著者情報
ジャーナル フリー

2005 年 21 巻 7 号 p. 851-854

詳細
抄録
A sensitive method for the preconcentration and determination of trace amounts of Co, Cu and Zn by energy-dispersive X-ray fluorescence spectrometry (EDXRF) has been developed. The method is based on the fact that 4-(2-pyridylazo)-resorcinol (PAR) loaded Dowex anion-exchange resin (PAR-resin) can effectively adsorb Co, Cu and Zn at pH 9.0 to form PAR-metal complexes. The detection limits for Co, Cu and Zn were 1.53, 0.31 and 0.21 ppb, respectively. The precisions for five replicate measurements of the three metals were 3.4, 2.7 and 2.1% RSD, and the calibration curves were linear up to 75 µg with correlation coefficients of 0.9975, 0.9980 and 0.9985, respectively. The method was successfully applied for the simultaneous determination of Co, Cu and Zn in seawater samples at ppb levels.