Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Optically Specific Detection of D- and L-Lactic Acids by a Flow-Injection Dual Biosensor System with On-line Microdialysis Sampling
Yoko NANJOTakayuki YANORyuzo HAYASHIToshio YAO
Author information
JOURNALS FREE ACCESS

2006 Volume 22 Issue 8 Pages 1135-1138

Details
Abstract

A flow-injection dual biosensor system with microdialysis sampling is proposed for the simultaneous determination of D-lactic and L-lactic acids. The dialysate from the microdialysis tube is delivered to a sample loop of the six-way autoinjector and then automatically injected into the flow-injection line with a dual enzyme electrode arranged in perpendicular to the flow direction. The dual enzyme electrode is constructed by hybridizing a poly(1,2-diaminobenzene) film into two sensing parts which respond selectively to D-lactic and L-lactic acids, respectively, without any cross-reactivity. The proposed flow-injection analysis method can be successfully applied to the simultaneous determination of D,L-lactic acids in alcoholic beverages.

Information related to the author
© 2006 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top