Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Construction and Evaluation of a Miniature Electron Ion Coincidence Analyzer Mounted on a Conflat Flange with an Outer Diameter of 114 mm
Takuhiro KAKIUCHIEiichi KOBAYASHIKoji K. OKUDAIRANarihiko FUJITAMasatoshi TANAKAKazuhiko MASE
著者情報
ジャーナル フリー

2008 年 24 巻 1 号 p. 87-92

詳細
抄録
We have developed a miniature electron ion coincidence (EICO) analyzer mounted on a conflat flange with an outer diameter of 114 mm. It consists of a cylindrical mirror analyzer (CMA), a time-of-flight ion mass spectrometer (TOF-MS), a commercially available linear motion feedthrough, and a tilt adjustment mechanism. Each sample surface was irradiated by synchrotron radiation, and the energies of emitted electrons were analyzed and detected by the CMA, while desorbed ions were collected by the TOF-MS in coincidence with the electrons. The performance of the EICO analyzer was tested by measuring the Auger-electron H+ photoion coincidence spectrum of condensed water at 4a1 ← O 1s resonance.
著者関連情報
© 2008 by The Japan Society for Analytical Chemistry
前の記事 次の記事