Simple and Sensitive Electroanalytical Method for the Determination of Ascorbic Acid in Urine Samples Using Measurements in an Aqueous Cationic Micellar Medium
Alaécio Pinheiro dos REISCésar Ricardo Teixeira TARLEYLucilene Dornelles MELLOLauro Tatsuo KUBOTA
著者情報
ジャーナル フリー

2008 年 24 巻 12 号 p. 1569-1574

詳細
抄録

An electroanalytical study for the amperometric determination of ascorbic acid (AA) in a human urine sample at a bare glassy carbon electrode using the CPC (cetylpyridinium chloride) surfactant is described. Under the optimized conditions of Epa, Ipa, CPC concentration and pH, the oxidation peak potential of the AA shifts towards a less positive potential, and the peak current increase in a significant way in the presence of the surfactant. Under the best conditions, the method provided a linear calibration curve to AA in a 0.1 mol L-1 phosphate buffer solution (pH 7.0) in the concentration range from 5.0 × 10-7 up to 4.3 × 10-4 mol L-1, with a detection limit of 2.0 × 10-7 mol L-1 and a high correlation coefficient (r = 0.9996). The recovery test presented values of 98 - 105%, suggesting a great potential of the proposed method for AA determination in complex samples, such as urine.