Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Heat-fixation Method Used in an Atomic Force Microscopy Study of Cell Morphology
Zhe-Xue LUZhi-Ling ZHANGWan-Liang SHIDai-Wen PANGZhi-Xiong XIEPing SHENG
著者情報
ジャーナル フリー

2008 年 24 巻 2 号 p. 257-260

詳細
抄録

In order to overcome the difficulties with existing methods for sample immobilization in imaging Halobacterium salinarum (H. salinarum) living in a highly salty medium by atomic force microscopy (AFM), a heat-fixation method was, for the first time, used to overcome existing problems in preparing samples for AFM. The effect on the cell morphology of the heat-fixation method was studied by MAC mode AFM, and was compared with the drop-and-dry and the polylysine-adhesion methods. It was found that the heat-fixation method can be successfully used for preparing Gram-negative and Gram-positive bacteria samples for AFM studies. Using this method, high-resolution AFM images of H. salinarum were obtained. Round protrusions on the cell surface and horn-like protrusions only at one pole of H. salinarum were observed.

著者関連情報
© 2008 by The Japan Society for Analytical Chemistry
前の記事 次の記事