Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Rapid Communications
Development of a Compact System for the Determination of Lead Using a Liquid Core Waveguide and a Blue Diode as a Light Source
Tomoki YABUTANIHonami KUROKAWAYuka HAYASHIHiromichi HIGUCHIYuki IWATAYuya HASHIZUMEShuji WADAYukako HAYASHIJunko MOTONAKA
著者情報
ジャーナル フリー

2010 年 26 巻 4 号 p. 407-409

詳細
抄録

A simple and compact system was developed for the on-site analysis of lead in environmental water samples. The system consisted of a custom-made blue diode as a light source, a liquid core waveguide (LCW) as spectrophotometric cell and a compact UV-visible spectrometer. It weighed within 1.5 kg in total. Lead was detected using a spectrophotometric indicator reagent (TPPS: 5,10,15,20-tetraphenyl-21H,23H-porphinetetrasulfonic acid, disulfuric acid, tetrahydrate) after masking interfering ions by coprecipitation with magnesium hydroxide. With an increase in the length of the LCW cell, the detection limit of lead was almost linearly enhanced, and the linear calibration range was shifted to a lower concentration range. The detection limit of lead by using 100 cm of a LCW cell was 2 nmol dm−3. The observed value (9.6 ± 0.4 μg dm−3) of lead in a river-water reference material (JSAC 0302-3) with this system was in good agreement with the certified value (9.9 ± 0.2 μg dm−3).

著者関連情報
© 2010 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top