Development and Validation of a RP-HPLC Method for the Determination of Zidovudine and Its Related Substances in Sustained-release Tablets
Jucimary V. dos SANTOSLuís A. E. Batista de CARVALHOM. Eugénia PINA
著者情報
ジャーナル フリー

2011 年 27 巻 3 号 p. 283

詳細
抄録
A reversed-phase high-performance liquid chromatography (RP-HPLC) method for the rapid and accurate quantification of zidovudine (AZT) in sustained release tablets during stability testing was developed. A Waters RP-18 XTerra™® column, using a water:methanol (80:20, v/v%) mobile phase at a flow rate of 1.0 ml min−1, and UV detection at 266 nm, was employed. The method of validation parameters indicate a linear range of between 40.0 to 220.0 μg ml−1 with an LOQ of 1.985 μg ml−1 and an LOD of 0.655 μg ml−1 for the analyte. The degradation products of AZT were isolated and characterized for the first time. There was a very little decline of antiviral by heat, and AZT did not completely degrade either by acid or alkaline hydrolysis. On the other hand, oxidation caused a higher degradation stress in the drug. Finally, the degradation products resulting from stress studies were not found to interfere with the detection of antiviral, which is an advantage of the presently proposed method.
著者関連情報
© 2011 by The Japan Society for Analytical Chemistry
前の記事 次の記事