Separation and Determination of Process-Related Impurities of Erlotinib Using Reverse-Phase HPLC with a Photo-Diode Array Detector
Chandrashekara KARUNAKARAUdupi APARNAVenkateshappa CHANDREGOWDAChandrasekara G. REDDY
著者情報
ジャーナル フリー
電子付録

2012 年 28 巻 3 号 p. 305

詳細
抄録
A simple and rapid reverse-phase high-performance liquid chromatographic (HPLC) method for the simultaneous separation and determination of erlotinib and its process-related impurities in bulk drugs has been developed. Five process-related impurities of erlotinib have been separated on an Inerstsil ODS-3V (C18) column and detected at 254 nm using a photo diode array (PDA). This HPLC method was successfully applied to the analysis of erlotinib bulk drug. The recoveries of erlotinib and process-related impurities were in the range of 92.86 – 106.23%, and found to be specific, precise and reliable for the determination of unreacted raw materials, intermediates in the reaction mixtures and bulk drugs.
著者関連情報
© 2012 by The Japan Society for Analytical Chemistry