Countercurrent Chromatographic Separation of Proteins Using an Eccentric Coiled Column with Synchronous and Nonsynchronous Type-J Planetary Motions
Kazufusa SHINOMIYAKazunori YOSHIDAKoji TOKURAEtsuhiro TSUKIDATEKazuhiro YANAGIDAIRAYoichiro ITO
Author information
JOURNAL FREE ACCESS

2015 Volume 31 Issue 3 Pages 211-218

Details
Abstract
Protein separation was performed using the high-speed countercurrent chromatograph (HSCCC) at both synchronous and nonsynchronous type-J planetary motions. The partition efficiency was evaluated with two different column configurations, eccentric coil and toroidal coil, on the separation of a set of stable protein samples including cytochrome C, myoglobin and lysozyme with a polymer phase system composed of 12.5% (w/w) polyethylene glycol 1000 and 12.5% (w/w) dibasic potassium phosphate. Better peak resolution was obtained by the eccentric coil than by the toroidal coil using either lower or upper phase as the mobile phase. The peak resolution was further improved using the eccentric coil by the nonsynchronous type-J planetary motion with the combination of 1066 rpm of column rotation and 1000 rpm of revolution.