Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Amperometric Determination of Ascorbic Acid on an Au Electrode Modified by a Composite Film of Poly(3,4-ethylenedioxythiophene) and Superconductive Carbon Black
Xinfei ZHOUKunyun HEYu WANGHaitao ZHENGShin-ichiro SUYE
著者情報
ジャーナル フリー

2015 年 31 巻 5 号 p. 429-436

詳細
抄録
A new composite film-modified electrode was prepared by the electropolymerisation of poly(3,4-ethylenedioxythiophene) (PEDOT) and superconductive carbon black (SCB) on a gold electrode. The PEDOT-SCB/Au electrode exhibited excellent ability towards the electrocatalytic oxidation of ascorbic acid, in terms of a 480 mV shift of the oxidation potential in the negative direction, and a dramatically enhanced oxidation current. Under the optimum conditions, the amperometric detection of ascorbic acid provided a wide linear detection range from 1.0 × 10−7 to 8.0 × 10−4 M, and a detection limit of 5.0 × 10−8 M (S/N = 2) as well as good reproducibility and stability.
Fullsize Image
著者関連情報
© 2015 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top