Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Effect of the sp2/sp3 Ratio in a Hybrid Nanocarbon Thin Film Electrode for Anodic Stripping Voltammetry Fabricated by Unbalanced Magnetron Sputtering Equipment
Hiroyuki YANAGISAWARyoji KURITATomoyuki KAMATAKyoko YOSHIOKADai KATOAyumi IWASAWATetsuya NAKAZATOMasaki TORIMURAOsamu NIWA
著者情報
ジャーナル フリー
電子付録

2015 年 31 巻 7 号 p. 635-641

詳細
抄録
The effect of the sp2/sp3 ratio in an unbalanced magnetron sputtered nanocarbon film electrode was studied for determining Cd2+ and Pb2+ by anodic stripping voltammetry (ASV). The signal-to-noise ratio in the ASV measurement improved as the sp3 concentration in the carbon film increased because the noise current decreased with the increasing sp3 concentration. The detection limits with a carbon film containing 50% sp3 were 0.25 and 1.0 μg L−1 for Cd2+ and Pb2+ with high repeatability (Cd: 4.6% and Pb: 6.4%, n = 3). For a real sample measurement, a pretreatment system combining a photooxidation reactor and a cation exchange column was used to eliminate the interference from EDTA and Cu2+, which forms a stable complex or alloy with Cd2+ and Pb2+. More than 99% of the interference was eliminated, and accurate signal currents for Cd2+ and Pb2+ were successfully obtained with the pretreatment system.
Fullsize Image
著者関連情報
© 2015 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top