Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Interference Reduction in Glucose Detection by Redox Potential Tuning: New Glucose Meter Development
Seong Je CHOChul-Ho CHOKwang Bok KIMMin-Hyoung LEEJae Hong KIMSuho LEEJaegeol CHOSuntae JUNGDong-Min KIMYoon-Bo SHIM
Author information
JOURNAL FREE ACCESS

2015 Volume 31 Issue 7 Pages 705-710

Details
Abstract

A new glucose meter was developed employing a novel disposable glucose sensor strip comprising a nicotinamide adenine dinucleotide-glucose dehydrogenase (NAD-GDH) and a mixture of Fe compounds as a mediator. An iron complex, 5-(2,5-di(thiophen-2-yl)-1H-pyrrol-1-yl)-1,10-phenanthroline iron(III) chloride (Fe-PhenTPy), was synthesized as a new mediator for the NAD-GDH system. Due to the high oxidation potential of the mediator, the detection potential was tuned to be more closely fitted toward the enzyme reaction potential, less than 400 mV (vs. Ag/AgCl), by mixing with an additional iron mediator. The impedance spectrometry for the enzyme sensor containing the mixed mediators showed an enhanced charge transfer property. In addition, a new cartridge-type glucose meter was manufactured using effective aligned-electrodes, which showed an enhanced response compared with conventional electrode alignment. The proposed glucose sensor resulted in a wide dynamic range in the concentration range of 30 – 500 mg dL−1 with a reduced interference effect and a good sensitivity of 0.57 μA mM−1.

  Fullsize Image
Content from these authors
© 2015 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top