Repeatability Assessment by ISO 11843-7 in Quantitative HPLC for Herbal Medicines
Liangmian CHENAkira KOTANIHideki HAKAMATARisa TSUTSUMIYuzuru HAYASHIZhimin WANGFumiyo KUSU
著者情報
ジャーナル フリー

2015 年 31 巻 9 号 p. 903-909

詳細
抄録

We have proposed an assessment methods to estimate the measurement relative standard deviation (RSD) of chromatographic peaks in quantitative HPLC for herbal medicines by the methodology of ISO 11843 Part 7 (ISO 11843-7:2012), which provides detection limits stochastically. In quantitative HPLC with UV detection (HPLC-UV) of Scutellaria Radix for the determination of baicalin, the measurement RSD of baicalin by ISO 11843-7:2012 stochastically was within a 95% confidence interval of the statistically obtained RSD by repetitive measurements (n = 6). Thus, our findings show that it is applicable for estimating of the repeatability of HPLC-UV for determining baicalin without repeated measurements. In addition, the allowable limit of the “System repeatability” in “Liquid Chromatography” regulated in a pharmacopoeia can be obtained by the present assessment method. Moreover, the present assessment method was also successfully applied to estimate the measurement RSDs of quantitative three-channel liquid chromatography with electrochemical detection (LC-3ECD) of Chrysanthemi Flos for determining caffeoylquinic acids and flavonoids. By the present repeatability assessment method, reliable measurement RSD was obtained stochastically, and the experimental time was remarkably reduced.

Fullsize Image
著者関連情報
© 2015 by The Japan Society for Analytical Chemistry