Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Time-Profile Measurement of an Emulsion Using Multiphoton Ionization Time-of-Flight Mass Spectrometry in Combination with a Microscope
Yosuke SHIMOTomohiro UCHIMURA
Author information
Keywords: MPI-TOFMS, O/W emulsion
JOURNAL FREE ACCESS

2016 Volume 32 Issue 10 Pages 1059-1063

Details
Abstract
Multiphoton ionization time-of-flight mass spectrometry was applied to the measurement of an oil-in-water emulsion that contained toluene as a dispersed phase. Before the measurement, the sample was sufficiently creamed, and then stirred for a short period of time for dispersion. As a result, several intense spikes appeared on the time profile constructed from the peak area for toluene. In the present study, an optical microscope was used to observe the capillary column for sample introduction, and small toluene droplets along with their aggregates were found in the images of the emulsion flowing through the capillary. The aggregates produced intense spikes comprised of multiple plots, which could be easily marked by applying a moving median filter. In the present study, droplets with minimum diameters as small as 2.9 μm, which corresponds to 13 fL, could be calculated as detectable spikes.
  Fullsize Image
Content from these authors
© 2016 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top