Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Electrochemical Detection of As(III) by a rGO/Fe3O4-modified Screen-Printed Carbon Electrode
Zhiyong XIEJuncui XUFazhi XIEShiquan XIONG
Author information
JOURNAL FREE ACCESS
Supplementary material

2016 Volume 32 Issue 10 Pages 1053-1058

Details
Abstract
A reduced graphene oxide (rGO)/Fe3O4 composites modified screen-printed carbon electrode (SPCE) was used to determinate As(III) in a HAc–NaAc buffer solution. The rGO/Fe3O4 composites were prepared by a simple and one-pot synthesis method, and characterized by X-ray diffraction (XRD), field emission scanning electron microscopy (SEM), FT-IR and Raman spectra. The electrochemical behaviors of the composite electrode were characterized by cycle voltammetry and electrochemical impedance spectroscopy. The experimental parameters, such as supporting electrolyte, solution pH, deposition potential, deposition time were optimized, respectively. The calibration curve for the detection of As(III) in the range of 2 to 20 ppb was I = –4.495 + 1.922C, with a coefficient of 0.994. The sensitivity was 1.922 μA/ppb, and with about twice Fe3O4 modified SPCE, a detection limit as low as 0.3 ppb was achieved. The proposed electrode was validated by analyzing the As(III) content in real water samples.
  Fullsize Image
Content from these authors
© 2016 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top