Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Perchlorate Selectivity of Anion Exchange Resins as Evaluated Using Ion-Selective Electrodes
Kenji YAMAMOTOShin’ya MITSUDANaomi OHTAKENatsuki MURASHIGESatoshi OHMUROAkio YUCHI
著者情報
ジャーナル フリー
電子付録

2017 年 33 巻 2 号 p. 159-163

詳細
抄録
The selectivity coefficients reported for perchlorate of the high selectivity on anion exchange resins (AXRs) have not been consistent with one another. Possible errors by the unique use of four parameters (concentrations of two anions in two phases) were experimentally verified. The concentrations of perchlorate buffered at low levels (10−6 – 10−4 mol L−1) by two forms of AXRs were successfully determined by potentiometry with a perchlorate ion-selective electrode. This gave reasonable coefficients. The coefficients for perchlorate on several AXRs were independent of the relative exchange (RE), in contrast to the previous reports. On the other hand, the coefficients for fluoride of the low selectivity that were examined for comparison decreased with an increase in RE, and the dependency was more remarkable for the resins of large exchange capacity.
Fullsize Image
著者関連情報
© 2017 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback