Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Fast and Simple Determination of 3-Aminopiperidine without Derivatization Using High Performance Liquid Chromatography–Charged Aerosol Detector with an Ion-Exchange/Reversed-Phase Mixed-mode Column
Shubo DONGZhengyu YANHanyue YANGZhen LONG
著者情報
ジャーナル フリー
電子付録

2017 年 33 巻 3 号 p. 293-298

詳細
抄録

A sensitive non-derivatization method for the determination of the highly polar compound 3-aminopiperidine was developed using a mixed-mode column combined with a charged aerosol detector (CAD). Chromatographic conditions, including the type of detector, separation mode, and mobile phase composition, were optimized to achieve high sensitivity towards and sufficient retention of 3-aminopiperidine. Compared to the precolumn derivatization UV method, the current method showed higher recovery and greater simplicify. High sensitivity (LOQ <2.73 μg mL−1) and good precision (RSD of peak area <2%) were also observed in the current method. Furthermore, the parameters such as buffer solution and column bleed that affected the sensitivity of the CAD were investigated. Finally, the current method was applied for the determination of 3-aminopiperidine in linagliptin samples. This is a new non-derivative for the determination of 3-aminopiperidine, and constitutes a novel application of the CAD for the quantitative analysis of highly polar basic compounds.

Fullsize Image
著者関連情報
© 2017 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top