Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Determination of Minor and Trace Metals in Aluminum and Aluminum Alloys by ICP-AES; Evaluation of the Uncertainty and Limit of Quantitation from Interlaboratory Testing
Michihisa UEMOTOMasanori MAKINOYuji OTAHiromi SAKAGUCHIYukari SHIMIZUKazuhiro SATO
Author information
JOURNAL FREE ACCESS
Supplementary material

2018 Volume 34 Issue 6 Pages 719-724

Details
Abstract

Minor and trace metals in aluminum and aluminum alloys have been determined by inductively coupled plasma atomic emission spectrometry (ICP-AES) as an interlaboratory testing toward standardization. The trueness of the measured data was successfully investigated to improve the analytical protocols, using certified reference materials of aluminum. Their precision could also be evaluated, feasible to estimate the uncertainties separately. The accuracy (trueness and precision) of the data were finally in good agreement with the certified values and assigned uncertainties. Repeated measurements of aluminum solutions with different concentrations of the analytes revealed the relative standard deviations of the measurements with concentrations, thus enabling their limits of quantitation. They differed separately and also showed slightly higher values with an aluminum matrix than those without one. In addition, the upper limit of the detectable concentration of silicon with simple acid digestion was estimated to be 0.03 % in the mass fraction.

Fullsize Image
Content from these authors
© 2018 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top