Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Refining Calibration Procedures of Circular Dichroism Spectrometer to Improve Usability
Kumiko SAKAI-KATOKohki YOSHIDATakashi OHGITAYuki TAKECHI-HARAYAYosuke DEMIZUHiroyuki SAITO
著者情報
ジャーナル フリー

2019 年 35 巻 11 号 p. 1275-1278

詳細
抄録

Circular dichroism (CD) is a technique used for conformational studies of peptides and proteins. We studied the specific calibration procedures of CD spectrometers based on procedures specified in the European Pharmacopoeia. We aimed to develop procedures to improve the usability of CD, in addition to reducing adverse effects on users’ health. The use of ethanol instead of 1,4-dioxane as the solvent for isoandrosterone was examined. Both solvents yielded the same maximum value of +3.3 for molar CD. We also studied a two-point calibration method using (1S)-(+)-ammonium 10-camphorsulfonate instead of (1S)-(+)-10-camphorsulfonic acid, which is a hygroscopic compound. Both compounds yielded similar results and the values for (1S)-(+)-ammonium 10-camphorsulfonate of 2.39 ± 0.04 and –4.92 ± 0.06 at 290.5 and 192.5 nm, respectively, were within the criteria defined in the European Pharmacopoeia. The inter-laboratory repeatability was also acceptable. These studies provide specific procedures for calibrating CD spectrometers for drug development.

Fullsize Image
著者関連情報
© 2019 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top