Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Application of Molecularly Imprinted Polymer-modified Potentiometric Sensor for Quantitative Determination of Histamine in Serum
Atsuko KONISHIShigehiko TAKEGAMIYoko IDOSAKITatsuya KITADE
Author information
JOURNAL FREE ACCESS
Supplementary material

2020 Volume 36 Issue 12 Pages 1561-1563

Details
Abstract

A molecularly imprinted polymer-modified potentiometric histamine (HIS) sensor was prepared and used for quantitative determination of HIS in bovine serum. The calibration curve using the potential responses measured in 1 × 10−3 mol L−1 phosphate buffer (pH 7.4) showed good linearity in the HIS concentration range of 3 × 10−4 to 1 × 10−2 mol L−1 (r = 0.92), with a detection limit of 1.6 × 10−4 mol L−1. In bovine serum samples, the HIS sensor showed good recovery values of 91 – 104%. Therefore, this HIS sensor successfully determined the HIS concentration in bovine serum samples.

Fullsize Image
Content from these authors
© 2020 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top