Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Semi-continuous Monitoring of Cr(VI) and Cr(III) during a Soil Extraction Process by Means of an Ion Transfer Device and Graphite Furnace Atomic Absorption Spectroscopy
Willy Cahya NUGRAHAHaruka NAGAIShin-Ichi OHIRAKei TODA
Author information
JOURNALS FREE ACCESS
Supplementary material

2020 Volume 36 Issue 5 Pages 617-620

Details
Abstract

Electrodialytic separation of Cr(VI) and Cr(III) followed by graphite furnace atomic absorption spectrometry for monitoring of soil extraction was studied. The sensitivity was improved by in-line purification of the solutions and bi-polar pulse cleaning. The detection limit for both Cr(VI) and Cr(III) was 0.01 μg L−1. The system was successfully used to monitor the concentration change during soil extraction with dual solution line filtration. The results demonstrate the difference in concentration changes with the different sources of Cr(VI).

Fullsize Image
Information related to the author
© 2020 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top