Notes
Preparation of N2-Ethyl-2′-deoxyguanosine-d4 as an Internal Standard for the Electrospray Ionization–Tandem Mass Spectrometric Determination of DNA Damage by Acetaldehyde
Yukihiro ESAKAHiromitsu ARUGASaki KUNISHIMATakuhei YAMAMOTOHiroya MURAKAMIYoshinari SAWAMAHironao SAJIKIBunji UNO
Author information
JOURNAL FREE ACCESS

2020 Volume 36 Issue 7 Pages 877-880

Details
Abstract

The deuteration of N2-ethyl-2′-deoxyguanosine (Et-dG), which is a DNA adduct generated from acetaldehyde, was studied by the addition reaction of acetaldehyde-d4 to 2′-deoxyguanosine (dG) in deuterium oxide (D2O), with the aim to obtain an isotope internal standard for the liquid chromatography/tandem mass spectrometry (LC/MS/MS) quantitation of Et-dG. The replacement of the dG C-8 hydrogen atom by a deuteron atom took place at 50°C in D2O and afforded a mixture of Et-dG-d4 and Et-dG-d5. Et-dG-d4, which was stable in aqueous solutions, was prepared by incubating the mixture in H2O at 60°C for 48 h. The calibration curve was obtained by multiple reaction monitoring (MRM) measurements using a hydrophilic interaction chromatography–electrospray ionization–tandem mass spectrometric (HILIC/ESI-MS/MS) system between the Et-dG concentration, ranging from 1.0 × 10−10 to 4.0 × 10−9 M in the sample solutions, and the relative peak areas of Et-dG (m/z: 296.1 → 180.1) to the value of Et-dG-d4 (m/z: 300.2 → 184.2), with an internal standard showing good linearity (R2 = 0.995, n = 5).