Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Preliminary Studies on Chemiluminescence Immunoassay for Human Chorionic Gonadotropin and Thyrotropin Using Acridinium Ester-Labeled Antibody
Naofumi SATOKamon SHIRAKAWAYouko KAKIHARAHiroshi MOCHIZUKITosinori KANAMORI
著者情報
ジャーナル フリー

1996 年 12 巻 6 号 p. 853-858

詳細
抄録
A highly sensitive chemiluminescence assay system for human chorionic gonadotropin (hCG) and thyroid stimulating hormone (TSH) was developed using a newly synthesized acridinium ester (AE) with a binding substitutional group at the N-10 position and having excellent chemiluminescence and stability. AE-labeled antibodies with an excellent chemiluminescence intensity were prepared after studying the labeling of anti-hCG antibodies with AE under different conditions. By studying the purification conditions for AE-labeled antibodies we produced AE-labeled antibodies having a low background value when assayed. When the detection limit is defined as the signal value+3SD at a concentration of zero for each substance measured, the detection limit for hCG is 0.01IU/l and that for TSH is 0.001μIU/ ml, demonstrating extremely high sensitivity. The labeled antibody′s stability is also excellent.
著者関連情報
© The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top