Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Determination of Trace Amounts of Silver in High-Purity Aluminum by Graphite-Furnace Atomic Absorption Spectrometry after Solvent Extraction and Micro-Volume Back-Extraction
Yoshio SHIJOEiji YOSHIMOTOTeruo KITAMURAHiroshi ONONobuo UEHARATokuo SHIMIZU
著者情報
ジャーナル フリー

1996 年 12 巻 6 号 p. 959-962

詳細
抄録
A method for determining trace amounts of silver in high-purity aluminum is described. Silver in an aluminum solution is extracted into xylene as its diethyldithiocarbamate (DDC) complex, and subsequently back-extracted into 200μl of nitric acid for a determination by graphite-furnace atomic absorption spectrometry (GFAAS). Silver can be completely extracted from an aluminum solution into xylene over a pH range of 2.5 to 4, and for a DDC concentration of over 4×10-3mol l-1. The resulting solution, back-extracted with nitric acid, is suitable for the determination of silver by GFAAS. The detection limit for silver was found to be 1.4ng g-1 when 0.2g of an aluminum sample is used. The application of this method to the trace analysis of silver in high-purity aluminum samples is demonstrated.
著者関連情報
© The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top