1991 Volume 7 Issue Supple Pages 1081-1084
Tributyltin(TBT) and triphenyltin compounds(TPT) in environmental and industrial waste waters were determined by gas chromatography with flame photometric detection using tripentyltin chloride(Pe3SnCI) as a surrogate compound. Recoveries of TBT and TPT from These water samples was 96-108%, respectively and the standard deviations(R.S.D) were 2.2-9.8%, 1.4 7.4%, respectively. The detection limits were 2ng/l, far TBT(as tributyltin chloride) and 6ng/l for TPT(as triphenyltin chloride).