An Electrochemical Biosensor Array for Rapid Detection of Alanine Aminotransferase and Aspartate Aminotransferase
Min-Jung SONGDong-Hwa YUNSuk-In HONG
Author information
JOURNAL FREE ACCESS Advance online publication

Article ID: 60043

Details
Abstract
An increment of alanine aminotransferase (ALT) or aspartate aminotransferase (AST) in human serum indicates an abnormal symptom of the liver. Hence, an electrochemical biosensor array that uses micro electro mechanical systems technology is required for rapid and integrated measurement of ALT/AST. Here we describe a biosensor array consisting of two glutamate sensors. It turned out that porous silicon layers formed on each working electrode were useful to increase the effective surface area. This biosensor array was constructed with platinum electrodes and a polydimethylsiloxane (PDMS) microchannel. Electrodes in sampling wells minimized a cross-interference effect and permitted multiple sampling by immobilization with glutamate oxidase using a silanization technique. The device sensitivities derived from semi-logarithmic plots were 0.145 μA/(U/l) for ALT and 0.463 μA/(U/l) for AST over a range of 1.3 U/l to 250 U/l. Hene, this ALT/AST biosensor array can be applied in diagnostic and home use.
Content from these authors

This article cannot obtain the latest cited-by information.

© 2009 by Japan Society for Bioscience, Biotechnology, and Agrochemistry